Tel: +86-769-22718086 | Alibaba
Product Description
1.1Name Calling:Thermal shock test chamber
1.2Type No:LJ102-5B
Battery thermal shock test box is suitable for testing all kinds of battery testIt is mainly used for furnace thermal test and heating test in battery safety performance test. Purpose: The final test requires that the battery cannot explode or catch fire. The battery thermal shock test chamber reaches the test temperature within a specified time at a certain heating rate and maintains it for a certain period of time. The hot air circulation system is adopted to reliably ensure the uniform distribution of working temperature.
GB/T 2900.11-1988 miner's lamp lithium ion battery safety performance test specification-thermal shock
YD 1268-2003 "Safety Requirements and Test Methods for Lithium Batteries and Chargers for Mobile Communication Handsets"-High Temperature Performance Test and Thermal Shock Test
SJ/T 11169-1998 Lithium battery-Heating test
QC/T 743-2006 Lithium-ion Batteries for Electric Vehicles-Heating
QC/T 744-2006 Nickel Metal Hydride Batteries for Electric Vehicles-Heating
UL 2054 Standard-Heating Test
GB/T 8897.4-2002 "primary batteries part 4 lithium battery safety requirements"-thermal abuse
SJ/T 11170-1998 Safety for Domestic and Commercial Batteries-Thermal Abuse
YDB 032-2009 backup lithium-ion battery pack for communication-thermal shock resistance, high temperature storage
Standard for performance testing of power batteries for electric vehiclesEV Li-ion-heating test
Performance Test Specification for High Power Lithium Ion Traction Battery for HEV-Heating Test
GB/Z 18333.1-Lithium-ion batteries for 2001 electric road vehicles-Heating test
QBT 2502-2000 General specification for lithium-ion batteries-High temperature test
GB/T 18287-2000 General specification for Li-ion batteries for cellular phones-High temperature performance, thermal shock
GB/T 2423.1-89 (IEC68-2-2)-High temperature test method
UL 1642 Standard-Heating Test
|
ItemObjective |
Name |
Inside Capacity |
|
Lord To Technology operation Reference Number |
Use power |
380V/50HZ(three-phase five-wireR,S,T,N,PE) |
|
Rated power |
9KW |
|
|
Studio Size |
Wide1000XDeep1000XHigh1000mm |
|
|
OutsideFormDimensions |
Wide2000*Deep1300*High1480MM |
|
|
Temperature range |
oftenTemperature 10℃-200℃ adjustable |
|
|
Temperature fluctuation |
≤ ±0.5℃ |
|
|
temperature uniformity |
≤ ±1.0℃ (no-load insulation) |
|
|
Temperature deviation |
±0.5℃ |
|
|
heating rate |
≤10℃/Min, the machine does not overshoot during the heating process. |
|
|
Wind transport mode |
Left and right forced air transport |
|
|
Box body knot Structure |
Shell Material |
1.50mmA3Cold plate bending welding production |
|
Surface treatment |
1.Surface grinding, polishing, phosphating and weld filling treatment |
|
|
2.Surface high gloss paint or corrugated enamel painting |
||
|
3.Color gray or user specified |
||
|
liner material |
1.0mmSUSThe stainless steel plate is bent and welded to keep the inner container clean and dust-free, rust-proof and corrosion-proof. The inner container and the outer shell are used as reinforcing ribs, and the heat insulation bridge is designed and manufactured to ensure the strength and heat insulation. |
|
|
Wind plate |
SUSThe stainless screen plate is fixed on the inner container by screws, which is convenient for disassembly and cleaning of the air duct. |
|
|
Insulation material |
unit weight120KGAluminum silicate wool, thickness100mm. |
|
|
Sealing material |
EType silicone sealing strip, temperature-resistant sealing, long service life. |
|
|
Vapor |
Diameter of upper part of box body120mmOne adjustable exhaust valve. |
|
|
Side with diameter90mmOne adjustable intake valve |
||
|
Insulation treatment |
The internal partition plate needs to be insulated., equipped with insulated bakelite. |
|
|
Box door |
Single door,Open from right to left. |
|
|
Caster |
The box body is equipped with four weighted universal limb wheels |
|
|
Plus Heat Department system |
Fan motor |
High temperature long shaft motor for oven380V/0.75KWOne |
|
wind wheel |
Diameter300mmcentrifugal wind wheel |
|
|
Heating pipe |
Flanged stainless steelUtype electric heating tube,with heat sink.The heating tube is wired outside the oven. |
|
|
Electricity Gas Control system |
Electric cabinet |
Right side of the box |
|
Control mode |
PLCTouch |
|
|
Temperature collection point |
middle position one |
|
|
Over-temperature protection thermostat |
TEEAPointing thermostat, the temperature exceeds its set temperature alarm and cut off the heater power to stop heating. |
|
|
Air switch |
Fuji air circuit breaker for power overload protection |
|
|
AC contactor |
Schneider AC contactor for heater power isolation, over temperature or when the time off. |
|
|
solid state relay |
High powerSSRsolid state relay, for heatingPIDOutput,Long contact-free mechanical life. |
|
|
Temperature sensor |
Thermocouple,K 0-200℃. |
|
|
Other electrical appliances |
Schneider Products |
|
|
Department system work Can |
Heating constant temperature |
Heating time and holding time can be set,Heating time to automatically cut off heating and alarm. |
|
Over-temperature alarm |
When the temperature in the box exceeds the temperature value set by the over-temperature protection thermostat, cut off the heating and give an alarm.. |
|
|
Three section lights and alarm |
The yellow indicator light is on before the oven heats up..Green light on during heating.At the end of the program or overtemperature, the red light comes on and sounds.. |
|
|
Power phase loss protection |
|
3.1Furnace shell surface temperature during operation≤Room temperature+10℃Outside temperature of furnace door≤Room temperature 15℃.
3.2When the equipment is manufactured, all welding parts must be firm and reliable, the welding seam must be free of slag, the welding seam must be polished and smooth, and the acute angle of the equipment appearance must be obtuse.
3.3After the manufacture of the equipment, the Supplier shall notify the Purchaser to send technical personnel to the Supplier for pre-acceptance.
3.4The supplier shall provide relevant data of the equipment (including the operation instruction document of the equipment manual and the list of wearing parts).
Previous Page
Previous Page
Next Page
Online Inquiry
