Reasons for slow cooling speed of high and low temperature test chamber
Source:
Release date: 2020-04-09
under normal circumstancesHigh and low temperature test chamberIt will cool to the corresponding value according to the set time and temperature, but due to the long use time or improper operation of the equipment, the temperature of the high and low temperature test chamber is likely to drop. In order to solve this problem, technical discussions will be conducted.
Check whether the refrigeration unit can be started during the operation period of the laboratory and whether the refrigeration compressor can be started during the operation period of the high and low temperature test chamber, indicating that everything is normal from the main switching power supply to the power supply circuit of each refrigeration compressor, and there is no problem in the circuit system.
There is no problem with the circuit system of the high and low temperature test chamber. Check the refrigeration unit again. First of all, check the exhaust pipe of the low temperature (R23) refrigeration compressor of the two refrigeration units in the high and low temperature test chamber, and the breathing working pressure is less than the standard value, and the breathing working pressure is in the exhaust condition, indicating that the refrigerant quantity of the main refrigeration unit is insufficient.
Touching the outlet and suction pipes of the R23 compressor of the main engine group by hand, it is found that the temperature of the exhaust pipe is not high and the temperature of the suction pipe is not low (no frost), which also indicates that the R23 refrigerant of the main engine group is insufficient.
The cause of the failure was not determined, and the cause of the failure was further confirmed in combination with the control process of the laboratory. The laboratory has two sets of refrigeration units. One is the main unit and the other is the auxiliary unit. When the cooling rate is high, the two units work at the same time. At the beginning of the insulation phase, both units are still working at the same time.
When the temperature starts to stabilize, the auxiliary unit stops working and the main unit maintains temperature stability. If the r23 of the host group leaks, the cooling effect of the host group will be small. Since the two units work at the same time during the cooling process, there is no temperature instability, and the indicated cooling rate will be reduced.
In the temperature maintenance phase of the high and low temperature test box, once the auxiliary unit stops working, the main unit has no cooling effect, and the air in the test box will slowly rise. When the temperature rises to a certain level, the control system will start the auxiliary unit to cool down, reduce the temperature to the set value, and then the auxiliary unit will stop working. If repeated, a fault will occur.
In summary, it has been confirmed that the cause of the failure of the high and low temperature test chamber is the leakage of refrigerant R23 from the low temperature (R23) unit of the main unit. Using the method of leak detector and soapy water, the leakage of the refrigeration system was detected, and it was found that the valve stem of the hot gas bypass solenoid valve had a crack of about 1cm. Replace the solenoid valve and refill the system with fluorine.