High and low temperature test chamberIt is suitable for the reliability test of high and low temperature of industrial products, and for the parts and materials of related products such as electronic and electrical engineering, automobile and motorcycle, aerospace, ship and aerospace, ship weapons, colleges and universities, scientific research units, etc. Under the condition of high and low temperature cross-charging cycle changes, test its various performance indicators. During the heating process of the high and low temperature test chamber, the circulating fan runs at high speed to accelerate the air convection in the working room, shorten the heating time and improve the work efficiency.
1. When the high and low temperature test chamber is in the whole process of temperature raising, the centrifugal fan of the circulation system runs at high speed to speed up the air convection in the room during work, reduce the temperature raising time and improve the working efficiency;
2. When the temperature in the room is close to the operating temperature during work, the instrument panel will automatically adjust the centrifugal fan of the circulation system from high-speed operation to low-speed operation, so that the gas in the room during work can gently heat convection, and reasonably increase the circulation system The service life of the centrifugal fan;
3. If full-automatic wind adjustment is not required, the main parameters of high and low speed gears can be adjusted to 999.9 according to the main parameter adjustment method, and the centrifugal fan of the circulation system will be in high-speed operation from beginning to end.
Self-tuning function of high and low temperature test chamber: if there is a large temperature overshoot or a large temperature fluctuation during the temperature control process, please start the self-tuning function according to the following operations
1, turn off the power switch, open the door, so that the high and low temperature test chamber natural cooling to ambient temperature;
2. Close the box door, turn on the power switch, and set the temperature to a common temperature value;
3. Adjust the self-tuning parameter to 1 according to the control parameter adjustment method;
4. Under normal working condition, press and hold the minus key for 5 seconds to enter the self-tuning state. At this time, the PV display shows ATU and flashes;
5. After self-tuning, the instrument automatically returns to normal working state;
6. When the high and low temperature test chamber is in the self-tuning state, pressing any key is invalid.