[Water-cooled xenon lamp aging test box] What are the requirements for water quality

Source:

Release date: 2019-04-15

The xenon lamp aging test chamber uses a xenon arc lamp that can simulate the full sunlight spectrum to reproduce the destructive light waves in different environments, which can provide corresponding environmental simulation and accelerated test for scientific research, product development and quality control.

Xenon lamp aging test chamber is divided into two forms; one is air-cooled; the other is water-cooled;If the water used in the water-cooled xenon lamp aging box does not circulate, the water consumption after a test is also very high. If it is to be recycled, the change of water quality and the temperature of the waterThenIt takes a lot of requests. The water-cooled xenon lamp aging test box for water quality requirements are very strict.ThatWhat are the requirements for water quality in the water-cooled xenon lamp aging test chamberWhat?What?

1. Deionized water must be used for the cooling water of the lamp tube, and the conductivity is below 5us/cm, because the triggering of xenon lamp will generate large current and high voltage. If water with high conductivity is used, the lamp tube will be damaged when the lamp tube is triggered..The temperature of the second inlet must beBelow 35 ℃, because the water temperature is too high will also lead to the service life of the xenon tube, so in the case of using circulating water, there must be a closed cold water system and filtration system to ensure that the water quality is secondary pollution due to cooling.

2. The water used for spraying water and humidification must also be distilled water. If ordinary tap water is used, white substances will appear on the surface of the sample, which cannot achieve the purpose of irradiation. This substance is called calcium and magnesium crystals produced after the impurities of water evaporate. Therefore, distilled water must be used to ensure the consistency of radiation and achieve the purpose of real testing.

3. There is also a point to be made about the voltage stabilization of the power supply, which is very important and is a key factor that directly leads to the use of water-cooled lamps.