[constant temperature and humidity box] how to solve the four main faults?
Source:
Release date: 2019-04-15
How to solve the four main faults of constant temperature and humidity box?
1. constant temperature and humidity box temperature and humidity display abnormal temperature
1. First check the ultra-wet protector in the control room to see if it is still at 150 ℃.
2. Secondly, check whether the pipeline in the laboratory has not been cleaned for too long, thus causing the water flow to be blocked. Then check whether there is water in the upper water tank. If there is no water, check whether there is water in the lower water tank.
3. Finally, check whether the solid-state relay of humidity is short-circuited. If there is no problem with the heater, we can use the AC voltage file of the electric meter to detect it.
4. Turn off the total power supply of the constant temperature and humidity box, open the drain valve under the heating barrel to a position parallel to the pipeline, drain all the water inside, clean the dirt inside, then pull out the silicone pipe, scrub in the water, and install it back in the opposite direction after maintenance. After the installation is completed, the drain valve under the humidifying bucket shall be closed to the position perpendicular to the pipeline, and then the main power supply behind the machine shall be turned on.
Frequent 2. at work during the day"Jumping" phenomenon, the cooling effect is normal at night
Analysis: The cooling effect is normal at night, indicating that the refrigeration system and circuit system are working normally. This shows that the constant temperature and humidity test chamber is caused by the external cause, and the reason is the poor heat dissipation of the condenser.
Check:The reasons for poor heat dissipation of the condenser are:
(1) Condenser indicates too dirty
(2) There is air in the refrigeration system
(3) The surrounding air does not flow
(4) The ambient temperature is too high
(5) Oil film on the inner surface of the condenser.
Processing results:The cooling effect of the test chamber is normal at night, which excludes the above reasons.(1), (2), (3) and (5). Move the test chamber to another place and it will no longer "jump". Therefore, it is determined that the constant temperature and humidity test chamber is close to the heat source and causes poor heat dissipation of the condenser.
Solution:Move the constant temperature and humidity test chamber to a place away from the heat source and ventilated to eliminate the fault.
How to solve the problem of overpressure in 3.
Most of the overpressure is caused by high ambient temperature. Confirm whether the room temperature is too high due to summer or lack of ventilation. Generally, the adaptive temperature of the constant temperature and humidity test chamber is5~28.
If the temperature of the room does not exceed the above mentioned, whether the equipment is too close to the wall or corner when placing the constant temperature and humidity test chamber, because this will lead to poor heat dissipation effect and alarm.
Another thing to consider is whether the test box has been cleaned since it was bought back. Generally, under the machine, that is, the refrigeration system, you can see the condensing fan, condenser and other large refrigeration component accessories, and whether the condensing fan and condenser are covered with dust. If so, please clean it up in time, because this small dust is preventing it from dissipating heat. At ordinary times, it is recommended to clean once a month or every quarter. When cleaning, you can use a vacuum cleaner or other auxiliary tools. Note that the cleaning must be operated in a shutdown state.
Four, hot and humid system failure
Possible performance: constant temperature and humidity box humidity display 100When the humidity is low, the humidity does not come down or the water tank is short of water. What should I do in these situations?
The solution to the slow rise of humidity or the failure to rise at all is as follows:
1. Check whether the humidity OUT on the controller has output or completely output
2. Check whether the SSR2 signal lamp is on. If it is on, you can measure the current with a clamp meter. If there is no current, it may be due to the following reasons:
a.FQ2 breaker tripped, cutting off the heating path;
B. Humidification solid state relay SSR is damaged, with signal input but not conducting;
c. Damaged humidifying pipe;