[high and low temperature test chamber] can be filled with samples for experiments?

Source:

Release date: 2019-04-22

First of all, the answer is no!

According to the standard requirements, the working chamber volume of the high and low temperature test chamber shall be at least the volume of the outer contour of the tested product.3~5 times, the reasons are as follows:

1. the tested product occupies the smooth channel after being placed in the box, the narrowing of the channel will lead to the increase of the air flow rate and accelerate the heat exchange between the air flow and the tested product. This is inconsistent with the reproduction of environmental conditions, because in the relevant standards, it is stipulated that the air flow rate around the test sample in the test chamber should not exceed1.7 m/s, in order to prevent the test sample and the surrounding atmosphere from generating unrealistic heat conduction, the average wind speed in the test chamber is 0.6~0.8 m/s under no load, not more than 1 m/s. When the space and area ratio specified by the above requirements are met, the wind speed of the flow field may increase by (50~100)%, and the average maximum wind speed is (1~1.7)m/s, meet the requirements of the standard. If the volume or windward cross-sectional area of the tested product is increased without restriction during the test, the airflow speed during the actual test will increase to exceed the maximum wind speed specified in the test standard, and the validity of the test results will be doubted.

The accuracy index of the environmental parameters in the 2. studio is the result of the detection in the no-load state, once the tested product is placed, the uniformity of the environmental parameters in the test chamber will have an impact, and the greater the space occupied by the tested product, the more serious the impact will be. The measured test data show that the temperature difference between the windward surface and the leeward surface in the flow field can reach3~8 ℃, severe can be more than 10 ℃. Therefore, it is necessary to meet the requirements specified in the standard as far as possible to ensure the uniformity of the environmental parameters around the tested product.

3., according to the principle of heat conduction, the temperature of the airflow near the box wall is usually different from the temperature of the center of the flow field.2~3 ℃, at the upper and lower limits of high and low temperatures, it may reach 5 ℃. The temperature of the box wall and the temperature of the flow field near the box wall are 2~3 ℃ (depending on the structure and material of the box wall). The greater the difference between the test temperature and the external atmospheric environment, the greater the temperature difference. Therefore, the space within the distance from the box wall (100~150mm) is not available.