Lijia Small Class: Common Faults in Low Temperature System [Cold and Hot Impact Box]

Source:

Release date: 2019-05-28

Hot and cold impact boxIt is one of the core products in the environmental testing industry. Its high temperature is generated by nickel-chromium alloy electric heating. The heating speed and the suppressible high temperature range are determined according to the power of the heating tube. The low temperature is realized by the refrigeration compressor. If there is a failure in the high temperature part, the heating tube or the over-temperature protector can be adjusted. If there is a failure in the low temperature system, it needs to be checked and judged by many parties. The following.Common Faults in Low Temperature System of Cold and Heat Impact Chamber:

 

1No cooling

First of all, determine whether the cooling button on the panel is pressed, and the ambient temperature of the test box should be ensured5℃~28°CBetween, look at the air-cooled condenser because of the long use time, the accumulated dust is too thick, clean up the dust on the condenser, if there is no above situation, you need to watch the reading of the refrigeration pressure gauge, the static pressure will generally be in8kg (or90psi,The innermost circle of the pressure gauge) is above, if it is below, there may be leakage of refrigerant, resulting in insufficient refrigerating capacity, and the refrigeration system needs to be overhauled..

 

2Slow cooling

   Determine whether the thermal load of the working chamber is too large or the blower does not rotate, or the test articles in the test box are stacked too much (under normal circumstances, it does not exceed 1/3 of the total volume of the inner box), causing the wind to fail to circulate and the temperature fluctuates greatly. Then check whether there is air leakage when the door is not closed properly, and check the breakdown of some solid state relays (such as heating2A solid state relay, there are1A breakdown) causes a part of the heating tube to be heated, offsetting a part of the refrigerating capacity, thus causing the illusion of slow cooling. If there is no such situation, it depends on whether the cooling of the test chamber is slow or cannot be reduced to the required temperature point at all, and the factor of adding refrigerant should be considered.

 

3,Hot and cold impact boxCompressor does not operate

    First, confirm whether the refrigeration button on the lower board is pressed, then open the distribution box, refer to the electrical schematic diagram and use a multimeter to measure the resistance value between the compressor power lines. If there is a resistance value (actually the resistance value of the compressor) and the resistance value is not much different, the compressor coil can be judged to be intact. If the resistance value is very different or not at all, there may be a short circuit or an open circuit in the coil.