Li Jia small class: [constant temperature and humidity test chamber] what is the reason for inaccurate humidity
Source:
Release date: 2019-06-10
Customers are usingConstant temperature and humidity test chamberthe time,More or less, you will encounter the problem that humidity cannot be increased or decreased.,Now, let's analyze this point and everyone together.
First, the constant temperature and humidity test chamber humidity and standard source is not accurate
1.The temperature sensor in the cabinet is damaged.
2.The measuring plate is not working properly.
3.The sensor is not well connected to the measuring board.
Second, the constant temperature and humidity box humidity display high solution
1.The wet bulb gauze is contaminated or incorrectly placed.
2.Sample or thermal load out of rated range.
3.Refrigeration system refrigerant leakage. Pipe blockage or solenoid valve does not work,Check the chiller current.
4.Frequent door opening,Evaporator frosting too thick.
5.Humidity transmitter bad.
Three, constant temperature and humidity test chamber humidity indication for a long time99.9% Solution
1.Wet ball gauze without water.
2.Wet bulb water solenoid valve broken.
3.Water purifier water supply is insufficient.
4.Humidity transmitter bad.
Four, constant temperature and humidity box humidity is low solution
1.The humidifier circuit wire protection switch trips or the AC contactor is damaged.
2.Humidifier working circuit solid state relay failure.
3.Insufficient water supply for humidification or solenoid valve of humidification system is broken.
4.Humidity transmitter bad.